61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
355
Sirona Dental Systems GmbH
9
Repair
Service Manual GALILEOS
9.12
Boards
båÖäáëÜ
15. Wait for approx. 1 minute. Then switch the unit back on.
The error message E6 15 05 (undefined system serial number) is
displayed.
16. Acknowledge the error message with the R key .
17. Error message E6 15 04 (undefined activation data) is displayed.
18. Acknowledge the error message with the R key .
19. Call the service menu [ → 210].
20. Call the service routine S008.3, check the serial number, and confirm
The unit serial number is located on the nameplate of the unit.
NOTICE! If the serial number is incorrect, exit the update process and
contact the Sirona Customer Service Center.
21. Switch off the unit.
22. Wait for approx. 1 minute. Then switch the unit back on.
The message
"No Key"
should no longer appear.
23. Call the service menu [ → 210].
24. Call the service routine S017 and perform the unit configuration (test
The board DX41 must be configured using the service routine
S017.9. In units with a serial number of 1080 and above, the
switching plate configuration of the swivel arm must be set or
checked using service routine S017.7.
Inform the customer of the configuration options of the software
status, for example, the acoustic exposure signal. Activate these
functions if they are required.
25. If the travel height of the unit has to be limited:
Set the travel height with service routine S018.2 [ → 272].
26. Perform a software update to the current software version [ → 61].
This updates all modules in accordance with the configuration.
The error message E1 11 20 (invalid unit calibration) is displayed.
27. Acknowledge the error message with the R key .
28. Perform a complete unit calibration [ → 155].
After a successful unit calibration has been performed, the error
message should no longer appear.
29. Call up the "Extended Details" via SiXABCon.
This generates an XML file (with the system parameters) which is
Filed in the PDATA/.../P2K_Config directory under the network name
of the unit.
● The process is completed.