Repair
Service Manual GALILEOS
9.12
Boards
båÖäáëÜ
14. Call the service routine S017 and perform the unit configuration (test
The board DX41 must be configured using the service routine
S017.9. In units with a serial number of 1080 and above, the
switching plate configuration of the swivel arm must be set or
checked using service routine S017.7.
Inform the customer of the configuration options of the software
status, for example, the welcome screen or acoustic exposure signal.
Activate these functions if they are required.
15. If the travel height of the unit has to be limited:
Set the travel height with service routine S018.2 [ → 272].
16. Perform a software update to the current software version [ → 61].
This updates all modules in accordance with the configuration.
17. Perform a complete unit calibration [ → 155].
After a successful unit calibration has been performed, the error
message should no longer appear.
18. Call up the "Extended Details" via SiXABCon.
This generates an XML file (with the system parameters) which is
Filed in the PDATA/.../P2K_Config directory under the network name
of the unit.
● The process is completed.