Maintenance
Service Manual GALILEOS
10.1
Calibrating the unit
båÖäáëÜ
10
Maintenance
Maintenance
IMPORTANT: Select the correct current/voltage type and adjust the
measuring range to match the expected readings.
10.1
Calibrating the unit
Calibrating the unit
Unit calibration is described in detail in the chapter entitled Adjusting/
calibrating the unit [ → 155] .
DANGER
Potentially lethal shock hazard!
It is essential to switch the unit off and to wait at least 1 minute, or 4
minutes if disconnecting the tube assembly (cable L3), before starting
the maintenance or taking off a cover panel!
CAUTION
Risk of electric shock!
Always switch the unit off before ...
...connecting a measuring instrument or
...carrying out continuity checks.
NOTICE
Risk of damage to boards
Please observe the usual precautionary measures for handling printed
circuit boards (ESD). Touch a ground point to discharge static electricity
before touching any boards.
NOTICE
Risk of damage to tube assembly
Keep to the prescribed cool-off periods if several exposures have to be
taken to check a measured value.