61 25 665 D3437
46
D3437
.
076.01.15
.
02 06.2012
3 Unit description
Sirona Dental Systems GmbH
3.2 Hardware
Service Manual GALILEOS
Scan time
approx. 14 s
Number of single exposures
200
Reconstruction:
Marking of focal spot:
Automatic exposure blocking:
The duration of automatic exposure
blocking (cooling period) depends on the
set kV/mA level and the actual exposure
time. Depending on the tube load, interval
times of 8 s to 300 s are automatically set
by the system.
Class I device
Degree of protection against electric
shock:
Type B device
Degree of protection against ingress of
water:
Ordinary equipment
(without protection against ingress of
water)
Year of manufacture:
(on the rating plate)
Mode of operation:
Continuous operation
Long-term power output:
100 W
Anode material:
Tungsten
Exposure parameters for determining
leakage radiation:
7mA/85 kV
Continuing current for leakage radiation
measurements:
0.14 mA
Transport and storage temperature:
Basic unit
-40°C – +70°C (-40°F – 158°F)
Detector
-30°C – +55°C (-22°F – 131°F)
Air humidity:
10% – 95% without condensation
Admissible operating temperature:
from +10°C to +35°C (50°F – 95°F)
Operating altitude:
≤ 3000 m
X-ray tube:
Toshiba DF-151R
or
Siemens SR 120/15/60
20XX