4 General operating procedures
Sirona Dental Systems GmbH
4.2 Updating the firmware
Service Manual GALILEOS
12. Click on the
"Show logfile"
button and use the log files to check
whether the update was successfully performed.
If it features entries such as "Update of DXxx failed!", please perform
the update again. Repeat this procedure as often as necessary until
the "failed messages" no longer appear.
13. Restart the unit now.
14. Use the software manager or the service routine S008.2 (see chapter
entitled "Unit software versions and compatibility") to check whether
all modules have been updated to the latest release of the program
(see chapter entitled "Using the update manager to check the
program releases".
15. Call up
"More details..."
via SiXABCon.
This generates an XML file (with the system parameters) which is
Filed in the PDATA/.../P2K_Config directory under the network name
of the unit.
NOTICE
Effectiveness of the software update
The unit must be restarted after every software update. The new DX11
version will not run until the unit has been rebooted (see also chapter
"Measures following replacement of boards [ → 347]").
Any errors with the consecutive numbers 01, 03, 04, 06 or 07 displayed
immediately following the software update may be ignored. If these
messages appear again after the unit is rebooted, perform
troubleshooting as described in the section entitled "Error
If anything conspicuous occurs in connection with unit handling on
completion of the software update and restart of the unit, please repeat
the software update as the first measure.