General operating procedures
Service Manual GALILEOS
4.2
Updating the firmware
båÖäáëÜ
4.2.2
Updating the Facescan firmware
Updating the Facescan firmware
There are two ways of updating the Facescan firmware:
● Option 1: Update via USB stick [ → 67]
The program data on the Facescan unit is completely overwritten.
● Option 2: Update via the network [ → 70]
The program data is transferred to the Facescan unit. Unlike in option
1, the unit configuration data is, however,
not
overwritten.
4.2.2.1
Option 1: Update using the Facescan USB stick
Option 1: Update using the Facescan USB stick
Opening Facescan
1. Unscrew the cover (A) from the Facescan.
2. Pull gray cable L78.4 from slot X2 of the FACESCAN modular board.
3. Unscrew the protective plate (T) from the Facescan unit.
ª
The FACESCAN modular board is visible.
4. Plug cable L78.4 again into slot X2 on the FACESCAN modular
board.