61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
83
Sirona Dental Systems GmbH
5
Messages
Service Manual GALILEOS
5.1
Help messages
båÖäáëÜ
IMPORTANT: The measures listed only clear help messages that result
from operator errors. If it is not possible to clear a message by taking the
measures listed, another type of error is the cause. In such cases, you
should run an error diagnosis [ → 88].
Help code
Description
Actions required
H3 01
"R button, move into starting position"
● Press the R key.
● Panoramic unit moves to starting position.
H3 20
"R button, confirm exposure data"
● Press the R key.
● Exposure data are confirmed.
H3 21
"Close the door"
● Close the door or check door contact.
H3 23
"Swivel pendant into end position"
● Move the swivel arm to its end position
(completely open or completely closed).
H3 24
The X-ray detector preparation is in progress. ● Wait until the X-ray detector is ready. This can
take up to 10 minutes.
H4 03
"Switch SIDEXIS to ready for exposure state"
● Make SIDEXIS XG ready for exposure.
H4 07
"SIDEXIS 3D Vorauswahl korrigieren"
Correct SIDEXIS XG 3D preselection.
H4 08
"SIDEXIS 3D Aufnahme wählen"
Correct SIDEXIS XG 3D preselection.
H4 20
"Get existing exposure"
IMPORTANT: Do not switch the system off until
the help message has disappeared.
● Get exposure with "Sirona Control Admin"
(see SIDEXIS XG
"Operator’s Manual"
(REF
59 62 134).