5 Messages
Sirona Dental Systems GmbH
5.5 List of error messages
Service Manual GALILEOS
E7 06 10
E7 06 12
E6 06 13
E2 06 20
E6 06 21
E6 06 22
Error code
Description
Actions required
see
E7 06 10
Module is stuck in bootloader stage. ● Check board DX6 (note LED states).
If the board remains in the bootloader
stage ...
● Repeat the software update.
● Replace the tube assembly.
Error code
Description
Actions required
see
E7 06 12
Unit is not ready for operation
This error may also occur in
connection with other causal error
messages! Please also observe the
causal error message! It appears
only after you acknowledge the first
error message.
● Check the CAN bus.
If this error occurs in combination with
other errors ...
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure and observe causal
error messages.
● Replace the tube assembly.
Error code
Description
Actions required
see
E6 06 13
Error when writing to
EEPROM.
Stored data may be lost.
● Acknowledge error and repeat
procedure.
If the error occurs repeatedly ...
● Replace the tube assembly.
Error code
Description
Actions required
see
E2 06 20
Overtemperature of single tank/
power pack
● Wait until the X-ray tube assembly has
cooled down.
● Check fan function by running service
routine S005.4; replace fan if
necessary.
● Check temperature sensor in single
tank by running service routine S005.5,
replace tube assembly if necessary.
Error code
Description
Actions required
see
E6 06 21
Hardware signal of release button
not detected.
● Check cable L5 (optical fiber), replace
if necessary.
● Replace board DX1.
● Replace the tube assembly.
Error code
Description
Actions required
see
E6 06 22
Broken temperature sensor
● Replace the tube assembly.