61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
91
Sirona Dental Systems GmbH
5
Messages
Service Manual GALILEOS
5.5
List of error messages
båÖäáëÜ
E3 06 23
E5 06 30
E5 06 31
E5 06 32
E1 06 40 GALILEOS
E1 06 41 GALILEOS
Error code
Description
Actions required
see
E3 06 23
Hardware signal of release button
applied during power-on.
● Check cable L5:
1. Switch off the unit.
2. Pull cable L5 off tube assembly.
3. Switch unit on.
4. Perform optical check of L5:
If light is visible ...
● Replace board DX1.
If no light is visible ...
● Replace the tube assembly.
Error code
Description
Actions required
see
E5 06 30
Total radiation time exceeded.
If a CAN bus error had been reported
before ...
● Check the CAN bus.
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
Error code
Description
Actions required
see
E5 06 31
Partial radiation time exceeded
If a CAN bus error had been reported
before...
● Check the CAN bus.
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
Error code
Description
Actions required
see
E5 06 32
Minimum preheating time not
observed.
If a CAN bus error had been reported
before...
● Check the CAN bus.
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
Error code
Description
Actions required
see
E1 06 40
Tolerance exceeded
VH nom.
● Replace the tube assembly.
Error code
Description
Actions required
see
E1 06 41
Tolerance exceeded kV nom.
● Replace the tube assembly.