61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
97
Sirona Dental Systems GmbH
5
Messages
Service Manual GALILEOS
5.5
List of error messages
båÖäáëÜ
E3 07 30/33/34/35/36 GALILEOS
E6 07 40
Error code
Description
Actions required
see
E3 07 30
Up/down keys pressed on power-
on.
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure and observe causal
error messages.
If the error occurs repeatedly ...
● Replace user interface with electronics
(DX7).
E3 07 33
Light localizer key pressed during
power-on.
E3 07 34
T key pressed during power-on.
E3 07 35
R key pressed during power-on.
E3 07 36
Touchscreen pressed during power-
on.
Error code
Description
Actions required
see
E7 07 40
No valid language set found.
Check selected language set by running
service routine S017.5, correct if
necessary.
Check whether selected language set is
already installed, perform software update
if necessary.
Please contact the Sirona Customer
Service Center (CSC) to find out whether a
bug fix by means of a software update is
possible and perform such an update if
necessary.
Summary of Contents for galileos