61 25 665 D3437
98
D3437
.
076.01.15
.
02 06.2012
5 Messages
Sirona Dental Systems GmbH
5.5 List of error messages
Service Manual GALILEOS
5.5.3
Location 10: System hardware
Location 10: System hardware
E7 10 02
E1 10 03
E1 10 04
E1 10 05: GALILEOS
E7 10 06 GALILEOS
Error code
Description
Actions required
see
E7 10 01
EEPROM cannot be written.
● Acknowledge error and repeat procedure.
If the error occurs repeatedly ...
● Replace board DX11.
Error code
Description
Actions required
see
E7 10 02
FPGA of DX1 is not addressable.
FPGA = Field Programmable Gate
Array
● Replace board DX1.
Error code
Description
Actions required
see
E1 10 03
The flash file system must be
formatted.
Occurs after replacement of board
DX11.
● Acknowledge error and repeat procedure.
The flash file system is formatted and error
code E110 04 is displayed.
Error code
Description
Actions required
see
E1 10 04
Flash file system formatting in
progress.
● Wait until the error code automatically
disappears (approx. 2 - 3 mins).
Error code
Description
Actions required
see
E1 10 05
Flash file system is not ready for
operation.
● Execute service routine S009.4 and
format flash file system.
The contents of the error memory are thus lost.
If the error occurs repeatedly ...
● Replace board DX11.
Error code
Description
Actions required
see
E7 10 06
Incompatible DX1-FPGA
(programmable logic component)
version for current operating mode.
● Check the hardware version of DX1 for
compatibility, replace board DX1 if
necessary.