Service Manual GALILEOS
5.5
List of error messages
båÖäáëÜ
5.5.4
Location 11: Power PC/Board DX11
Location 11: Power PC/Board DX11
E6 11 02: GALILEOS
E6 11 03: GALILEOS
E7 11 04
E6 11 05: GALILEOS
E7 11 07
Error code
Description
Actions required
see
E6 11 01
Program sequence error.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
● Acknowledge error and repeat procedure.
If the error occurs again ...
● Reset the entire calibration of the unit and
readjust the unit.
● Replace board DX11.
Error code
Description
Actions required
see
E6 11 02
Watchdog error
● Acknowledge error and repeat procedure.
If the error occurs repeatedly ...
● Replace board DX11.
Error code
Description
Actions required
see
E6 11 03
Operating system/resource error.
● Acknowledge error and repeat procedure.
If the error occurs repeatedly ...
● Replace board DX11.
Error code
Description
Actions required
see
E7 11 04
Implausible data in EEPROM.
● Check the unit configuration via service
routines S017 and S018 and reconfigure if
necessary.
● Check calibration with diaphragm test
exposures.
S.
If the calibration is not OK ...
● Recalibrate the unit.
If the calibration is OK ...
● Make the individual unit settings again
(e.g. programming of the patient symbol
keys; see Operating Instructions).
Error code
Description
Actions required
see
E6 11 05
RAM allocation failed.
● Replace board DX11.
Error code
Description
Actions required
see
E7 11 07
Unknown or invalid definition of unit
class.
Occurs during first power-on after
replacement of board DX6 or DX11.
● Take the action required after replacing a
board.