5 Messages
Sirona Dental Systems GmbH
5.5 List of error messages
Service Manual GALILEOS
5.5.7
Location 15: Configuration, update
Location 15: Configuration, update
E7 15 03
E6 15 04
*) OI = Operating instructions
E6 15 05
E6 15 10
Error code
Description
Actions required
see
E7 15 01
Wrong memory modules.
If a DRAM memory module is plugged into
board DX11 ...
● Replace memory module or DX11.
If no DRAM memory module is plugged into
board DX11 ...
● Replace board DX11.
Error code
Description
Actions required
see
E7 15 03
Wrong software constellation of
modules.
● Check the software versions of the unit (on
the info screen or by running service
routing S008.2), and run or repeat
software update or downgrade if
necessary.
Error code
Description
Actions required
see
E6 15 04
Product activation keys invalid or
not available.
Occurs after replacement of tube
assembly (DX6) or board DX11 and
possibly after software updates.
See also the section titled Measures
● Enter release key.
see OI*
Error code
Description
Actions required
see
E6 15 05
Unit serial number invalid or not
available.
Occurs during first power-on after
replacement of board DX6 or DX11.
See also the section titled Measures
● Run service routine S008.3 and confirm or
enter the unit serial number on the unit.
Error code
Description
Actions required
see
E6 15 10
Update file for module is corrupt.
● Obtain latest update file from the Sirona
Customer Service Center (CSC) or the
Sirona home page and perform software
update.