61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
107
Sirona Dental Systems GmbH
5
Messages
Service Manual GALILEOS
5.5
List of error messages
båÖäáëÜ
5.5.8
Location 41: Media interface card
Location 41: Media interface card
E6 41 02
E6 41 03
E6 41 04
E6 41 05
Error code
Description
Actions required
see
E6 41 01
General module initialization error.
● If the error is a software error known to the
Sirona Customer Service Center (CSC), a
software update (bug fix) must be
performed.
● Replace board DX41.
Error code
Description
Actions required
see
E6 41 02
Invalid system data or uninitialized
module storage data
● If the error is a software error known to the
Sirona Customer Service Center (CSC), a
software update (bug fix) must be
performed.
● Acknowledge error and repeat procedure.
If the error occurs again ...
● Replace board DX41.
Error code
Description
Actions required
see
E6 41 03
Invalid commanding or control data.
This error may also occur in
connection with other causal error
messages! Please also observe the
causal error message! It appears
only after you acknowledge the first
error message.
● Check the CAN bus.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
Error code
Description
Actions required
see
E6 41 04
Data transfer error or dialog error to
module (master side)
● Check the CAN bus.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
Error code
Description
Actions required
see
E6 41 05
Data transfer error or dialog error to
bootloader of module
Only occurs in connection with
software update
● Repeat the software update.
● Check the CAN bus.
● Replace board DX41.