Adjusting/calibrating the unit
Service Manual GALILEOS
7.3
Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
7.3.1.2
"Type 3" diaphragm
"Type 3" diaphragm
Adjusting the "diaphragm open"
diaphragm setting
✔
The
"Service functions"
✔
The
"Diaphragm"
tab is selected.
✔
The
"Collimator open"
is selectable in the
"Image acquisition"
menu
area.
1. Set the rotary knob on the tube assembly to the "open diaphragm"
position.
2. Click on the
"Collimator open"
button.
ª
SIDEXIS XG makes the unit ready for exposure [ → 162].
ª
Service routine S030.5 is displayed on the control panel.
3. Take an exposure (85 kV/21 mAs) [ → 162].
4.
For "GALILEOS Comfort":
Evaluate the image.