Adjusting/calibrating the unit
Service Manual GALILEOS
7.3
Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
7.3.8
Geometry calibration
Geometry calibration
✔
The
"Service functions"
1. Click the
"Geometry"
tab.
ª
The corresponding tab card is selected.
2. Read off the serial number of the geometric phantom from the ID
label of the phantom and enter it in the
"Seriennr. Kalbrierphantom"
text box of the
"Geometry"
tab card.
3. Insert the geometric phantom in the bite block holder of the
4. For Facescan: Check to make sure that the room lighting is switched
on. Switch it on if necessary.
5. Click on the
"Image acquisition"
button.
ª
SIDEXIS XG makes the unit ready for exposure [ → 162].
ª
Service routine S010.13 is displayed on the control panel.
6. Press the R key to move the unit back to the starting position.
IMPORTANT
For an installed and configured Facescan:
The face scanner is automatically calibrated during the geometry
calibration of GALILEOS. In this case, make sure that the normal room
lighting is switched on during the calibration process. The room does not
have to be darkened during calibration.