61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
191
Sirona Dental Systems GmbH
7
Adjusting/calibrating the unit
Service Manual GALILEOS
7.5
Checking and adjusting the touchscreen
båÖäáëÜ
7.5
Checking and adjusting the touchscreen
Checking and adjusting the touchscreen
1. Switch the unit on.
ª
Once the unit software has started up, the start screen is
displayed on the touchscreen.
2. As soon as the start screen appears, press the 3 keys light localizer,
T, and R simultaneously (see (A)).
IMPORTANT: These keys must be pressed while the unit is booting
and the start screen is still displayed.
ª
The first adjustment screen appears on the display.
ª
You can abort the adjustment procedure at any time by pressing
the
"Abort"
button.
3. Touch the top left corner of the screen.
ª
The second adjustment screen appears on the display.
4. Touch the bottom right corner of the screen.
ª
The third adjustment screen appears on the display.
5. Touch the center of the green square in the top right corner of the
screen.
IMPORTANT: Touch the green squares as close to the center as
possible so that the black dots that then appear will be positioned
exactly in the center of the squares.
ª
The fourth adjustment screen appears on the display.
A