61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
193
Sirona Dental Systems GmbH
7
Adjusting/calibrating the unit
Service Manual GALILEOS
7.6
Mechanical adjustments
båÖäáëÜ
7.6
Mechanical adjustments
Mechanical adjustments
7.6.1
Ring center adjustment
Ring center adjustment
1. Remove the "arm cover".
2.
Move the ring center to the left or right:
NOTICE! Do not undo the screws completely! Make sure that spring
(B) does not pop out. This spring has a defined prestress!
Loosen the four screws (A) slightly.
3. Correct the position carefully and then retighten the screws.
NOTICE
Perform this adjustment only if the measured values are out of
A
A
A
A
B