Service
Service Manual GALILEOS
8.5
S005: General X-ray tube assembly service
båÖäáëÜ
8.5.2
S005: Test step 4
S005: Test step 4
Fan test
* Factory setting
1. Call service routine S005.4 [ → 213].
2. Use the arrow keys (A) in selection field 1 to select the code "01" (see
table).
3. Confirm your selection by pressing the R key (B).
ª
The fan starts up.
4. Check the fan for running noise.
5. Exit the service routine [ → 220].
Upon exiting the service routine, the fan is automatically switched off
again.
Selection field
Code
Function
1
00
Fan off*
01
Fan on