Service
Service Manual GALILEOS
8.6
S007: Error logging memory
båÖäáëÜ
8.6
S007: Error logging memory
S007: Error logging memory
* SR=service routine, ** SHZ=security access
8.6.1
S007: Test step 1
S007: Test step 1
Display error logging memory
In addition to service routine S007.1, you can also use the extended detail
query in SiXABCon to check the error logging memory.
Title only for GALILEOS
Easypad touchscreen
* Factory setting
1. Call service routine S007.1 [ → 213].
2. Use the patient symbol keys (B, C, D) to select the step width for
scrolling between the error events (see table).
ª
The selected patient symbol key lights up.
SR*
SHZ**
Function
S007
Error logging memory
S007.1
No
Display error logging memory
S007.2
Yes
Clearing error logging memory
S007.5
No
Enabling CAN bus logging in the web interface
Symbol on the control
panel
Status
Function
Patient symbol key 1 (B) is selected Step width for scrolling
between error events = 1*
Patient symbol key 2 (C) is selected Step width for scrolling
between error events = 10
Patient symbol key 3 (D) is selected Step width for scrolling
between error events = 100
Selection field
Selection/display
1
Error event
2
Error code for the selected event
3
Date and time of the selected
error event