8
Service
Service Manual GALILEOS
8.8
S009: Flash file system
båÖäáëÜ
8.8.3
S009: Test step 7
S009: Test step 7
Save/restore DX89 data
1. Call service routine S009.7 [ → 213].
ª
After you select this service routine, the following can be
displayed in selection field 1:
DX89 ⇒ DX11:
Data is transferred from DX89 to DX11, memory key (A)
(Easypad) or LED above the memory key (A) (Multipad) is lit up.
DX11 ⇒ DX89:
Data is transferred from DX11 to DX89, memory key (A)
(Easypad) or LED above the memory key (A) (Multipad) is lit up.
"---":
Data on both boards (DX11 and DX89) is valid or the data
transfer is not possible, all keys (Easypad) or LEDs (Multipad) are
not lit.
Only one practical direction of data transfer is offered at any one
time. If both locations contain valid data,
"---" is displayed.