8 Service
Sirona Dental Systems GmbH
8.11 S017: Configuration service
Service Manual GALILEOS
8.11
S017: Configuration service
S017: Configuration service
* SR=service routine, ** SHZ=security access
8.11.1
S017: Test step 2
S017: Test step 2
Configuring the hardware version
* Factory setting
SR*
SA**
Function
S017
Unit configuration
S017.2
Yes
Confirming the unit version
S017.3
Yes
Enter the country group code
S017.4
(for "GALILEOS Comfort")
Yes
Select a language
S017.5
(for "GALILEOS Comfort")
Yes
Select a language set
S017.6
Yes
Enable/disable the remote control
S017.7
Yes
Configure the switching plate for the swivel arm
S017.9
Yes
Enable/disable operation with board DX41
S017.13
only for
("GALILEOS Comfort")
Yes
Enable/disable the welcome screen
S017.14
(for "GALILEOS Comfort")
Yes
Enable/disable certain lines of the welcome screen
S017.15
Yes
Activate/deactivate the acoustic signal for end of exposure
S017.25
Yes
Select the diaphragm type
Selection field
Code
Function
1
0100
CB*
0500
CB incl. Facescan