8.11 S017: Configuration service
Service Manual GALILEOS
8.11.2
S017: Test step 3
S017: Test step 3
Enter the country group code
* Factory setting
S017_3 GALILEOS
1. Call service routine S017.3 [ → 213].
ª
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
Selection field
Code
Function
1
00
Worldwide*
01
Asia
02
US