8 Service
Sirona Dental Systems GmbH
8.11 S017: Configuration service
Service Manual GALILEOS
8.11.11
S017: Test step 25
S017: Test step 25
Select the diaphragm type
* Factory setting
1. Call service routine S017.25. [ → 213]
ª
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
Selection field
Code
Function
1
00
Type 1 diaphragm
("GALILEOS Compact")
01
Type 1/Type 2 diaphragm
("GALILEOS Comfort")
02
Type 3 diaphragm
("GALILEOS Compact")
03
Type 3 diaphragm
("GALILEOS Comfort")*