8.12 S018: Service for height adjustment
Service Manual GALILEOS
8.12
S018: Service for height adjustment
S018: Service for height adjustment
* SR=service routine, ** SHZ=security access
8.12.1
S018: Test step 2
S018: Test step 2
Set the maximum travel height
S018_2 GALILEOS
1. In user mode, move the unit to the required maximum travel height by
pressing UP/DOWN keys (A).
2. Call service routine S018.2 [ → 213].
ª
Once the service routine has been selected, the current height
position is displayed in selection field 1.
ª
Memory key (B) (Easypad) or the LED above Memory key (B)
(Multipad) lights up.
SR*
SA**
Function
S018
Service for height adjustment
S018.2
No
Set the maximum travel height
S018.3
No
Undo the maximum travel height setting
S018.4
No
Check the height adjustment sensor system