61 25 665 D3437
66
D3437
.
076.01.15
.
02 06.2012
4 General operating procedures
Sirona Dental Systems GmbH
4.2 Updating the firmware
Service Manual GALILEOS
4.2.1.2
Check program releases
Check program releases
You can use the right-hand window of the software manager to check
which modules are connected to the unit and what their latest program
release is.
NOTE: From unit serial number 8500 (for "GALILEOS Comfort") or 48500
(for "GALILEOS Compact") and above, a DX11_FPGA version is also
displayed in addition to this list.
Modules which are connected and whose program release corresponds
to the latest main software version (see chapter entitled "" are identified
by a continuous green bar.
Modules which the system does not recognize are identified by a broken
red bar.
If the actual status of the module cannot be polled for the update, the
actual SW version will be displayed as = V00:00.
If a module has a hardware incompatibility to the program status to be
programmed or the software version on the module is newer than the one
in the update file, this will be indicated by a red triangle with an
exclamation mark.
If the version of the selected update file is lower than the current software
version of the unit, then there will be no display in the right window. The
downgrade required in this case is possible only via
"Main version"
mode.