076.01.15
.
02 06.2012
4 General operating procedures
Sirona Dental Systems GmbH
4.3 Configuring the unit
Service Manual GALILEOS
4.3.2.3
Resetting the Facescan configuration to factory default settings
Resetting the Facescan configuration to factory default settings
Opening the web dialog
1. In SIDEXIS Manager, start the
"Facescan Configuration"
application.
2. Click on the
"Facescanner in browser"
button.
ª
A password dialog box opens.
3. In the field
"User"
enter
"service"
.
4. In the field
"Password"
enter
"sirona"
.
ª
The
"Facescan Device Service "
web dialog opens.
Selecting the service dialog
Selecting the service dialog
➢ In the menu bar, select the menu item
"SERVICE"
.
ª
The
"Facescan Service Functions"
window opens.
Reset
1. Press the
"Settings Reset"
button.
ª
A reset dialog opens.
2. Press the
"Settings Reset"
button.
ª
The configuration is reset to the factory settings.
NOTICE
Complete loss of user data
When the unit is reset to factory settings, all user data is overwritten,
including white balance and calibration data.
The network settings are, however, retained.