61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
81
Sirona Dental Systems GmbH
4
General operating procedures
Service Manual GALILEOS
4.5
Using demo mode – operation without radiation release
båÖäáëÜ
4.5.2
Switching off demo mode
Switching off demo mode
1. Switch off the unit.
2. Remove the cover of the tube assembly.
3. Loosen the screws (A) and remove the cover plate (B).
4. Set the dip switch S2 (DX6) to position 1.
5. Connect cable L5 (XRAY) to connector J6 (DX6).
ª
Radiation release is now once again possible.
6. Switch on the unit and check the mode on the info screen.
Demo mode: ON means that: Demo mode is switched on (radiation
release is not possible)
Demo mode: OFF means that: Demo mode is switched off
(radiography, X-ray radiation are possible!)
7. Switch the unit off again and reattach the cover plate (B) and the tube
assembly covers by following the dismantling procedure in reverse
order.
DANGER
Potentially lethal shock hazard!
It is essential to switch off the unit and to wait at least another 4 minutes
before taking off the covers of the X-ray tube assembly.