61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
85
Sirona Dental Systems GmbH
5
Messages
Service Manual GALILEOS
5.4
Error messages
båÖäáëÜ
5.4
Error messages
Error messages
GALILEOS error messages
The error messages are displayed as error codes (Ex yy zz) on the
Easypad touchscreen (GALILEOS Comfort) or on the Multipad display
(GALILEOS Compact) as well as on the remote control display (if there is
one).
The codes provide you with error type, error location and troubleshooting
information.
5.4.1
Error code: Ex yy zz
Error code: Ex yy zz
The error messages are encoded according to the following pattern:
5.4.2
Ex - Error type
Ex - Error type
Identifier x is intended to help you reach a decision quickly on how to
proceed with the corresponding error.
Ex
Error type
“Troubleshooting” classification for the user
yy
Location
Module, subsystem or logical function unit
zz
Consecutive
number
Identification of error
x
Description
Error group
Actions required
1
System warning
System message
This error group includes all errors
that indicate still acceptable
tolerance variations, or messages
about states which do not directly
affect system operation.
● Acknowledge the error message.
If the error occurs again ...
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure
If the error occurs again ...
● Run an error diagnosis [ → 88].
2
Errors caused by
system overload
This error group includes states
that indicate temporary
overtemperatures or similar, for
example. The cause of the error
disappears automatically after a
certain waiting time.
● Acknowledge the error message.
● Repeat the procedure step after a certain
waiting time.
If the error occurs again ...
● Extend the waiting time.
If the error occurs again ...