62 14 972 D 3507
30
D 3507
.
031.01.07
.
02 11.2010
Electrical connection
Sirona Dental Systems GmbH
Connecting the arm cable
Installation Instructions HELIODENT
PLUS
6
Electrical connection
6.1
Connecting the arm cable
➢ Connect the arm cable to terminal strip X100 on the opposite side so
that the lead color codes match.
Lead colors (from top to bottom on terminal strip X100):
● 1 : blue (V)
● 2 : pink (W)
● 3 : gray/brown (negative)
● 4 : white (heating-)
● 5 : yellow ()
● 6 : green (kVact)
➢ Secure the arm cable with a cable tie for strain relief.
X100
X100