Sirona Dental Systems GmbH
Electrical connection
Installation Instructions HELIODENT
PLUS
Connecting the release and the door contact
båÖäáëÜ
6.4.3
Installation option 1
empty
Configuration
Installation option 1
Release in the treatment room without remote control
● Release
– Manual release S3
empty
Installation
1. Connect manual release S3 (coiled cable) to X401 on board DX1 at
connectors X401.1 and X401.4.
2. Secure the coiled cable with the shielding clamp (B).
3. Check the setting of switch S1 on PC board DX4. Switch S1 must be
set to "
external
". (see section entitled "Selection of the release
4. Check the setting of switch S2 on PC board DX4. Switch S2 must be
set to "
ON
" (see section "Service routine S15" [ ➙ 69]).
1
S3
X401.1
X401.2
X401.3
X401.4
L3
1
S3
DX1
NOTICE
The shielding of the coiled cable must be in flat contact with the shielding
clamp (B).