Sirona Dental Systems GmbH
Electrical connection
Installation Instructions HELIODENT
PLUS
Connecting the release and the door contact
båÖäáëÜ
6.4.6
Installation option 4
empty
Configuration
Installation option 4
Release outside of the X-ray room with remote control
● Release
– Manual release S3
or
– Remote control release key S9
empty
Prerequisites
● The remote control has been installed (see section "Installing the
remote control/remote timer [ ➙ 23]").
● Only with use of the manual release S3 (coiled cable):
– The holder for manual release S3 was attached to the cover of
the remote control. See section "Installation on the protective
cover of the remote control [ ➙ 29]".
empty
Installation of cable L9 (on DX1)
1. Connect cable L9 to connectors X401.1 and X401.4 on PC board
DX1.
2. Secure cable L9 with the shielding clamp (B).
4
L9
S3
S9
NOTICE
Installation prerequisites
Use of the remote control is permissible only if the yellow X-Ray LED is
visible to the operating personnel during radiation release.
L9
4
X401.1
X401.2
X401.3
X401.4
S3/S9
DX1
X401
B
L9
NOTICE
The shielding of cable L9 must be in flat contact with the shielding clamp
(B).