Starting up
Installation Instructions HELIODENT
PLUS
Functional test
båÖäáëÜ
test
1. Switch the measurement voltage source on.
2. Read the voltage drop at the voltmeter and adjust the measuring
range if necessary.
3. Switch the measurement voltage source off.
Calculation
The measured value is the exact equivalent of the leakage current.
Limit value
The measured value must not exceed 0.55mA.
7.1.4
Checking the X-ray tube assembly
1. Switch on the building installation or the emergency circuit to which
the HELIODENT
PLUS
was connected.
2. Switch the unit on.
3. Monitor the self-test.
4. Select the shortest radiation time (0.01 - 0.05) and 60kV.
5. Release radiation 3 times.
WARNING
Potentially lethal shock hazard
Do not touch the unit while measuring the leakage current!
Measuring range
Leakage current
mV
µA
V
mA
NOTICE
Always document your measuring results for comparative
measurements performed later.
➢ Enter the measured values in the maintenance certificate.
➢ Describe the measuring instrument used (to ensure reproducibility).
WARNING
X-ray radiation!