Installing the manual block clamp
Installation Instructions for Manual Block Clamp CEREC MC XL, inLab MC XL
Final work
bå
Öä
áë
Ü
3. Remove the calibration phantom by grasping it by its narrow surfaces (B).
4. Insert the milling instruments in the DC motor manually. Tighten the cor
-
responding chuck with the torque wrench until a clicking sound can be
heard.
5. Close the milling chamber door.
Ç
The dialog box for selecting the milling instruments then appears.
6. Select the inserted milling instruments and confirm by clicking the
"Start"
button in the window.
Ç
The DC motors move to their starting positions.
The
"Calibration succeeded"
dialog box appears.
7. Confirm with the
"OK"
button.
empty
Supplement to the Operating Instructions
Hand over the document "Supplement to the Operating Instructions" to the
customer. Instruct him to store this document together with the Operating In
-
structions.
WARNING
Sharp edges on scanner!
The scanner on the left motor mount has sharp edges which could cause
personal injury.
Be careful not to brush against the scanner with your hand while removing
the calibration phantom.
NOTICE
Store the calibration tools in a safe place
Store the calibration pins and the calibration phantom in a safe place (e.g.
in the supplied drawer insert). For this purpose, replace the existing left dra
-
wer insert with the one supplied.