Using the manual block clamp
Installation Instructions for Manual Block Clamp CEREC MC XL, inLab MC XL
bå
Öä
áë
Ü
empty
Blocks with 6 mm diameter block holders
Fastening the block
1. Insert the adapter sleeve (A) into the block clamp.
2. Insert the block (B) into the block adapter sleeve.
3. Clamp the ceramic block securely with the ball pressure screw (C). Use
the block clamp tool with torque wrench for this purpose.
Ç
The block is pressed laterally against the contact surface of the block
clamp and simultaneously pulled in axially. The plate of the block hol
-
der thus rests on the block clamp.
empty
Removing the adapter sleeve
1. Loosen the ball pressure screw.
2. Place the block clamp tool in the inner groove (D) and pull out the adapter
sleeve.
NOTICE
Fasten the block tightly
If the block is not tightened sufficiently, this may result in falsification of the
milling result and fracturing of its ceramic material.
Fasten the block tightly using the block clamp tool with torque wrench un
-
til you hear a click.
Check to make sure that the block is seated very firmly.
NOTICE
Insert the adapter sleeve
The slot at the bottom end of the adapter sleeve must lie above the radial
pin of the block fastener in order to be inserted fully.
The hole for the ball pressure screw is then automatically in the correct po
-
sition, i.e. coincides with the threaded hole in the block clamp.