11 Commissioning, demo mode and repacking
Installation Manual ORTHOPHOS SL
11.7 Configuring the unit
båÖäáëÜ
11.7
Configuring the unit
Room and travel height
1. Check the room height before you raise the unit using the UP/DOWN
keys.
2. If the room height is less than 2.27 m (89 3/8") or 2.30 m (90 1/2") for
installation with the floor stand, you must limit the maximum travel
height using service routine S018.2.
Unit variants
For unit orders with the ceph arm, the unit version must be configured
through service routine S017.2 [ → 138].
Remote control
All units ordered with or without remote control are preconfigured for this
version at the factory. You should still check the unit configuration using
service routine S017.6.
Occlusal bite block
All units ordered with or without occlusal bite block are preconfigured for
this version at the factory. You should still check the unit configuration
using service routine S017.18.
Set up the occlusal bite block in accordance with customer requirements
using service routines S018.7 through S018.10.