11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.8 Setting up the X-ray component
Installation Manual ORTHOPHOS SL
11.8
Setting up the X-ray component
Since it is addressable via the network, the X-ray component can in
principle be activated for X-ray image acquisition by any of the PCs
connected to the network.
IMPORTANT: In order to avoid IP address conflicts, you should never
operate several networkable X-ray components in the network using the
same IP address. Each X-ray component must be assigned a unique IP
address and the name of the X-ray component must also be unique.
Factory setting of the TCP/IP address of the unit:
192.168.15.240 (subnet mask: 255.255.255.0)
NOTE: The network configuration can be set via service routine S037.
Checking existing IP addresses
To find out whether an IP address already exists in the network, enter the
"PING" function in the input prompt (e.g. in MS-Windows “CMD.EXE”).
1. Switch on
all
network devices (computers, printers, X-ray
components) which have been operated to date.
2. Invoke the input prompt (DOS window) from a network computer.
3. At the prompt, enter "ping" followed by the address to be checked and
then press the Enter key.
Example: “ping 192.168.15.13”
If a network device responds, then this address has already been
assigned.