11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.11 Using demo mode – operation without radiation release
Installation Manual ORTHOPHOS SL
1. Switch the unit on and move it to its packing height by actuating the
up/down keys on the control panel.
Bite block height = 965 mm (displayed as height on the control
panel)
Bottom edge of slide cover = 702 mm
2. Move the diaphragm and the sensor unit to the packing
position [ → 161] via the service routines S021.3 and S021.4.
3. Turn the rotation unit to the packing position (see section
"Delivery [ → 28]" / "ORTHOPHOS SL [ → 30]".
4. IMPORTANT: Attach the tube assembly guard (supplied in the
package) to the X-ray tube assembly and secure the packing position
of the tube assembly with the tension belt (supplied in the package).
5. Attach the transport safety device.
The transport safety device is attached in reverse sequence to how it
6. Pack the panoramic X-ray unit
(Packing condition - see section Delivery [ → 28]).
Packing the cephalometer
1. Call service routine S034.6 [ → 133].
After calling the service routine, an inactive progress indicator is
displayed in selection field 1. Selection field 2 shows the
characters "FFFF".
2. Touch the T key (A).
The cephalometer moves to the packing position. The procedure
is visualized by an active progress indicator in selection field 1.
When the cephalometer has reached its packing position, "0000"
is displayed in selection field 2.
3. Exit the service routine [ → 136].