11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.12 Service routines for commissioning and repacking
Installation Manual ORTHOPHOS SL
11.12.3.1
S017: Test step 2
Configuring the hardware version
* factory settings
1. Call service routine S017.2.
Once the service routine has been selected, the current selected
unit variant is displayed in selection field 1.
2. Use arrow keys (A) to select the required unit variants in selection
field 1 (see table).
Once the unit variant has been selected, the Memory key (B)
lights up.
3. Save the setting.
4. Exit the service routine.
Selection
field
Unit variants
Function
1
PanDCS__3D_______ Panoramic digital, 2D/3D-Sensor
PanDCS__3D_CephL Panoramic digital, 2D/3D-Sensor,
Ceph left digital
PanDCS__3D_CephR Panoramic digital, 2D/3D-Sensor,
Ceph right digital
PanDCS____________ Panoramic digital, 2D-Sensor
PanDCS______CephL Panoramic digital, 2D-Sensor,
Ceph left digital
PanDCS______CephR Panoramic digital, 2D-Sensor,
Ceph right digital