64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
149
Sirona Dental Systems GmbH
11 Commissioning, demo mode and repacking
Installation Manual ORTHOPHOS SL
11.12 Service routines for commissioning and repacking
båÖäáëÜ
11.12.4.1
S018: Test step 2
NOTE: Limiting the max. travel height is only possible from a position
value of >1500.
Set the maximum travel height
1. Move the unit to the required maximum travel height by pressing the
Up/Down keys in the user mode on the control panel.
2. Call service routine S018.2.
Once the service routine has been selected, the current height
position is displayed in selection field 1.
The Memory key (B) lights up.
3. To save the maximum travel height, touch the Memory key (B) (R key
(C) lights up) followed by the R key (C).
4. *For greater clarity the unit is shown in the adjacent image without a
cover.
Set the mechanical limit stop on the unit:
Loosen the nut (D) and move mechanical limit stop (E) for the limit
switch until it engages. Tighten nut (D) again.
The next time the UP key is pressed, the unit stops 10 mm below the
limit switch.
5. Exit the service routine.
D
E
*
Summary of Contents for ORTHOPHOS SL 2D
Page 2: ......