11 Commissioning, demo mode and repacking
Sirona Dental Systems GmbH
11.12 Service routines for commissioning and repacking
Installation Manual ORTHOPHOS SL
Adjusting and saving the PWM index
values
1. Call service routine S018.10.
The PWM index for upward travel "U" (up) is displayed in
selection field 1.
The PWM index for downward travel "D" (down) is displayed in
selection field 2.
The display of patient symbol key 1 (G) indicates whether or not
the occlusal bite block is connected (see table).
2. Select the desired index values using the arrow keys in selection
fields 1 and 2 (see table).
3. To save the index values, first press the Memory key (the R key lights
up) and then the R key.
4. Exit the service routine.