02 03.2016
12 Startup, measurements and tests (for USA/Canada only)
Sirona Dental Systems GmbH
12.5 Checking the tube voltage
Installation Manual ORTHOPHOS SL
12.5
Checking the tube voltage
Preparing the measurement
1. Attach the Mult-O-Meter sensor in the middle of the sensor unit (DCS-
side).
2. Set the main switch (A) to I (see also Operating Instructions).
3. Wait for approx. 1 minute.
4. Press the R key.
The unit moves to its starting position.
Performing measurements
1. Select the P1 program (see Operating Instructions).
2. Set the kV/mA level 62 kV/8 mA (see Operating Instructions).
3. Make ready for exposure in SIDEXIS 4 (see Operating Instructions).
4. CAUTION! Activating the release button triggers X-rays.
Start the exposure by pressing the release button. Hold down the
release button until image acquisition is completed and the acoustic
signal that indicates the end of the exposure can be heard.
Analyzing measurements
➢ Read the voltage values on the display of the Mult-O-Meter.
If the measured values
do not fall within the permissible tolerance
range
, replace the
tube assembly
(see Service Manual).
If the measured values are within the permissible tolerance
range, finalize the measurement.
Concluding the measurement
➢ Switch the unit on via switch (A) (see also Operating Instructions).
A
IMPORTANT
The measured tube voltage must correspond with the tube voltage set
of 62kV. The permissible tolerance is ± 10 %.