13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.1 General information on unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.1.2.2.4
Automatic adjustment
2D adjustment/calibration
During the automatic 2D adjustment via the calibration menu, default
values for adjustment are displayed in the menu's text boxes after every
exposure. Repeat automatic adjustment until the default values in the text
boxes are equal to "zero". (This means that adjustment has been
successful.)
If the values do not converge toward "zero" despite repeating the
adjustment procedure several times, calculate the adjustment values
manually by measuring the exposure with the toolbar measuring ruler and
overwrite the default values in the menu. This procedure is described in
the following sections.
3D adjustment/calibration
During the automatic 3D adjustment via the calibration menu, whether the
adjustment was successful or must be repeated again is displayed in the
message window in plain text.