64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
215
Sirona Dental Systems GmbH
13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2
Adjustment and calibration via the calibration
menu
Only for Germany
13.2.1
2D adjustment/calibration
13.2.1.1
Pan sensor adjustment
IMPORTANT
After every adjustment and calibration of the unit, the reference values
for the constancy measurement must be recalculated and entered in the
"Test results" form, "Reference value" column (see technical document
"ORTHOPHOS SL 2D/3D Acceptance test DIN 6868 Part 151 / Part
161"
(REF 65 61 059).
Summary of Contents for ORTHOPHOS SL 2D
Page 2: ......