64 95 142 D3632
220
D3632
.
031.01.02
.
02 03.2016
13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
3. Overwrite the adjustment values for S1, S2 and S3 in the text boxes
of the
"Sensor adjustment"
menu with the measured values.
For information about the shifting direction (input of +/- sign in the
menu), see the chapter entitled "Shifting direction of the exposed
NOTICE! Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter