64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
225
Sirona Dental Systems GmbH
13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2.1.2.2
Manual adjustment: Pan - Diaphragm
The manual adjustment procedure is similar to the one for automatic
adjustment. The only difference is that the default adjustment values
calculated automatically in the
"Diaphragm adjustment"
menu are
overwritten by adjustment values calculated manually.
1. Start the
"Diaphragm adjustment"
adjustment procedure as
described in the chapter entitled "Automatic adjustment: Pan
diaphragm (standard) [ → 221]".
2. Once you have taken the exposure, measure distances Sx and Sy
with the toolbar measuring ruler [ → 197].
Tip: For better measurement, you can also edit [ → 197] the exposure
using the toolbar functions.
Summary of Contents for ORTHOPHOS SL 2D
Page 2: ......