13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
4. Repeat the procedure starting at Step 2 until the values in the menu's
text boxes are equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT: If automatic adjustment fails to produce values equal
to zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 230].
5. If adjustment is OK (the adjustment values in the text boxes are equal
to "zero"), save the values. [ → 209]
The status indicator that prefixes the
"Symmetrie adjustment"
element is checked.
6. Remove the needle phantom from the bite block holder of the unit.
7. Continue with the next calibration point. [ → 231]
A
Adjustment OK
● The shadow of the center needle,
the needle image, and the
auxiliary lines must be coincident
and located one behind the other.
A tolerance (offset of needle from
the central auxiliary line) of ± 0.75
mm is admissible.
● Distance A1 must be 88.6 ± 1 mm.
● Distances A2 must be identical,
each being 44.3 ± 0.5 mm.
B
● A surrounding white border must
be present.
C
Adjustment not OK