13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
6. Repeat the procedure from point 5 for all other kV/mA combinations.
After the last exposure has been taken, a message in the action
window indicates whether calibration has been successful and
the status indicator at the start of the element
"Sensor calibration
(DCS)"
in the structure tree responds accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
7. Continue with the next stage of the adjustment procedure [ → 233].