13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
1. In the structure tree, under
"Ceph"
, click on the
"Adjustment of Ceph
secondary diaphragm"
element (S010.5).
The
"Adjustment of Ceph secondary diaphragm"
menu is
displayed in the action area. The menu supports precision
adjustment and coarse adjustment (precision adjustment is pre-
selected). Perform a precision adjustment first. In most cases,
previous coarse adjustment is not necessary. [ → 203]
2. Establish readiness for exposure [ → 209].
3. Create an exposure (80kV / 14 mA; 0.6 s) [ → 209].
The adjustment values S1, S2 and S3 are automatically
determined from the exposure and entered in the text boxes of
the
"Adjustment of Ceph secondary diaphragm"
menu.
The exposure is displayed in the exposure window.