13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
3. Set the diaphragm tilt via screw (C) (Lx in mm) and the diaphragm
height via screw (D) (Ly in mm).
Screw (C):
Turn counterclockwise = adjusts the diaphragm to the right
Turn clockwise = adjusts the diaphragm to the left
Screw (D):
Turn counterclockwise = adjusts the diaphragm downward
Turn clockwise = adjusts the diaphragm upward
To measure the adjustment, use a gauge.
The above-mentioned directions of rotation for the screws apply to
both left-arm and right-arm versions.
4. Retighten screws (B) firmly.
5. Once adjustment is complete, put the secondary diaphragm cover
back again and fix it in place with the screw (A).
C
D
Lx
Ly